With a comprehensive repair and calibration package for precision instruments, Noavaran Payesh Company offers repair and calibration services for all types of vibration meters, alignment systems, and thermography cameras. As you are aware, precision instrumentation generally requires annual calibration. Furthermore, if you lack confidence in the data displayed by your equipment, please send the device to us for calibration.
Scenarios indicating suspicious data include non-repeatability. For instance, during the alignment process, when performing an alignment check and rotating the device to the 9, 12, and 3 o’clock positions (in the direction of motor rotation), the data may not be repeatable on the second or third attempt, or the data may not correspond with dial indicator measurements (specifically the Reverse Dial method). In such conditions, calibration is certainly required.
Additionally, regarding thermal cameras or thermometers, if the device does not display the temperature of boiling water as 100°C (considering the camera’s tolerance, typically ±2°C), calibration is required.
If you believe your device requires calibration or specific repairs, you may send the device to the company’s headquarters.
For immediate coordination, please contact +98 912 071 4820.
**Devices repairable at Noavaran Payesh Company include:**
* All products from LGP (Turkey), including Pocket Balancer, Pocket Route, Vibra Align, etc.
* All products from Erbessd Instruments (Mexico), including Wiser 3x, MX 10, MX20, MX30, etc.
* All products from VMI (Sweden), including Easy-Viber, X5, etc.
* All products from Easy-Laser (Sweden), including XT440, E950, E960, etc.
* Various vibration meter sensors, including CTC, etc.
* All thermography cameras, including brands such as Fluke, LGP, Testo, FLIR, etc.